Technical parameters:
Product Size | 153mm (L) × 162mm (W) × 227mm (H) Allowable error ± 5% |
Product Weight | 1.7kg |
Mode Selection | drip rate mode, volume mode, time mode |
Infusion Set Specification | special or standard disposable infusion set |
History recording function, maximum 2000 history can be stored. | |
Infusion Rate Range | ml/h Mode: (1 to 1200) ml/h, 1ml/h increment per step Drop/min Mode: (1 to 266) drops/min |
Infusion Rate Accuracy | ml/h mode: within ±5% (after calibration) |
Drop/min Mode | within ±3% (after calibration) |
Mechanical Accuracy of Infusion Pumps | within ±2% |
Rapid Infusion Rate | 100 ml/h~1000 ml/h (100 ml/h per step increment) |
KVO Speed | 1ml/h~5ml/h (each increment is 1ml/h) |
Infusion volume | 1ml~9999ml |
Cumulative Infusion Volume | 0ml~9999.9ml |
Blocking alarm threshold | Height: 800mmHg±200mmHg (106.7kPa±26.7kPa) |
Medium: 500mmHg±100mmHg (66.7kPa±13.3kPa) | |
Low: 300mmHg±100mmHg (40.0kPa±13.3kPa) | |
Alarm | Infusion completed, pipeline blocked, door opened, air bubbles, battery low, etc. |
Power Supply | 220VAC , 50Hz; built-in battery: 11.1V rechargeable lithium battery pack, capacity ≥1600mAh; after the new battery is fully charged, the infusion pump can work continuously for more than 4 hours at a rate of 25ml/h. |
Bubble Detector | ultrasonic detection mode, detection sensitivity ≥25μL. |
Fuse | specification T2AL/250VAC, quantity 2, installed in infusion pumps |
IP Rating | IPX2 splash-proof equipment |
Safety Class | Class I, internal power supply, CF type continuous operation equipment |
Operating environment conditions: a) Ambient temperature range | +5℃~+40℃ |
b)Relative Humidity | 20~90% |
c) Atmospheric Pressure Range | 86.0KPa~106.0 KPa |
Transportation and storage conditions: a) Ambient temperature range | -20 ℃ ~ +55 ℃ |
b) Relative humidity | ≤93% |